Sterilization Instructions


Sterilization Instructions


The Cautery Handpiece and Cautery Electrode may be autoclaved according to the instructions below.


Separate the electrode from the Handpiece prior to autoclaving and use aseptic procedures when handling and sterilizing both items.


 
  1. Wipe and remove all debris present on the Handpiece/Electrode.
  2. Place the Handpiece/Electrode in an autoclavable pouch or wrapped cloth.
  3. Steam sterilize for a period of 20 minutes at 250°F(121°C) (Do not flast autoclave or exceed 250°F). Consult manufacturer's recommendations, or the AAMI recommended practice for Steam Sterilization Using the Wrapped Method.
  4. Allow the Handpiece/Electrode to cool for 30 minutes prior to use.
  5. Inspect the Handpiece/Electrode after sterilization for damage. If the Handpiece/Electrode has deteriorated or cracked, or severe discoloration or damage has occurred, discard and replace.

CAUTION: Do not use liquid sterilizing solutions with the Geiger TCU or accessories, as the potential for material damage exists.


 